SS869 8/12/08 Rev BApply power to calibrator with AC adapter or charge battery.Turn on power switch located in the top right corner. Power is on when
10 5.3 USER INTERFACE PANEL The user interface panel includes the pressure and electrical connections to the unit under test. (See Diagrams 5A &am
46 11 5.4 GETTING STARTED 1. Open calibrator top. 2. Turn on calibrator power (power switch on right side of the calibrator). 3. Once calibrator
126.0 POCKET PC INTERFACE SCREENS The Pocket PC Interface is tab-based, split into 6 major sections at the bottom of the screen. 6.1 APPLICATION DE
4412.0 SPECIFICATION DERIVATIONSTotal Error Band(Worst case) over atemperature range of-10°F to +130°FNon Linearity: ±0.1% FSHysteresis: ±0.05% FSNon
14 6.3 TEST: General Screen Instructions (Diagram 8): 1. Tap the TEST tab at the bottom of the page 2. Select desired UUT ID from the drop down b
4210.0 GLOSSARY OF TERMS Atmospheric Pressure - Pressure of the atmosphere at the Earth’s surface. NIST standard atmospheric pressure = 1.01325 bar
16 6.33 TEST: Dither Control Page (Manual Pressure Incrementing) Screen Instructions (Diagram 10): Test Point Indicator: Displays present test poi
40 Once a UUT ID is selected, the test results list will update to display only the test results from that UUT. (See Diagram 36) The test results fo
18 6.4 PROFILE: General Setting up Test Profiles Database (Diagram 11): 1. Profiles are pressure settings used during testing. 2. Program has cap
38 New Button - This button opens a dialog box to create a unique profile. Profile ID must be entered into the ID box on the dialog. 1. To copy a pr
20 6.5 UUT (UNIT UNDER TEST) SETUP: General Screen Instructions (Diagram 13): UUT setup screeen is used for inputting UUT profiles into the UUT da
36 Pocket PC Database - This list shows what test profiles are currently on the Pocket PC. After editing any profile, said profile will automaticall
22 UUT SUPPLEMENTAL PAGE 1. Enter the UUT Part Number. 2. Enter the UUT Model. 3. Enter the UUT Manufacturer. 4. Enter the required pressure Settlin
349.0 MICRO-CAL MANAGER INTERFACE The MICRO-CAL MANAGER INTERFACE provides a central location for all data within a given facility, database manageme
24 6.51 UUT (UNIT UNDER TEST) SETUP: Dial Gauge Differences from General (Diagram 13): -Output fields hidden. -Excitation field hidden. 6.52 UUT
32Electrical LinesUnit Under Test(Model 260)Pneumatic LinesDiagram 22_ + EXC24VDCSIG(+)SIG(-)COMCurrent InternalExcitationEXC COM SIG
26 6.7 SYSTEM Screen Instructions (Diagram 17): Data and system values: Software Version - Version of software on Pocket PC. Firmware Version -
27Diagram 17Default System SettingEng. Units SettingSoftware VersionFirmware VersionSerial NumberFilter LengthExcitationBattery LevelLeak Test System
28 EDIT DEFAULTS PAGE Display trend graph on UUT warm-up box will make a graph appear when the DoTest Button is pressed on test page.The graph will
3Table of Contents1.0 SAFETY INSTRUCTIONS ... 52.0 INTRODUCTION ...
49.0 MICRO-CAL MANAGER INTERFACE 9.1 Setup for Pocket PC Connection ... 34 9.2 Graphical User Interface Setup ...
51.0 SAFETY INSTRUCTIONS Use the calibrator only as specified in the manual, otherwise the protec-tion provided may be impaired. Do not operate
62.0 INTRODUCTION DOCUMENTING PROCESS CALIBRATOR: The Setra Micro-Cal Model 869 is a pressure generating, very low pres-sure documenting calibrator
50 73.0 STANDARD EQUIPMENT BREAKOUTTHE ITEMS BELOW ARE INCLUDED IN YOUR CALIBRATOR:Diagram 2AC Power AdapterPneumatic Fittings & TubingElectronic
84.0 OPERATING FEATURES 4.1 THE POCKET PC PERFORMS THE FOLLOWING MAJOR FUNCTIONS • User interface for setups, display and storage. • Data trans
4813.0 SPARE PARTS & ACCESSORIES Part # Item 269701 269 terminal block 269425-02 security key 869902 accessory bag 869904-xx elec