7.0 AGENCY ELECTRICAL STANDARDS
This device falls into CSA “Pollution Degree 2” for PCB insulation and
CSA “Installation Category 2”.
The SRPM meets the following requirements:
CSA Standard C22.2 No 0-M 91:
General Requirements - Canadian
electrical code, Part 1
CAN/CSA C22.2 No. 0.4-04:
Bonding of Electrical Equipment
CAN/CSA-C22.2 No. 61010-1-04: Safety requirements for electrical
equipment for measurement, control and laboratory Use Part-1: General
Requirements
ANSI/UL61010-1 (Second Edition): Safety requirements for electrical
equipment for measurement, control and laboratory use Part 1: General
Requirements.