4.0 OPERATION For most accurate pressure measurement, allow the Model 730 to warm up for at least 15 minutes. After installation, periodically check

Page 3 - Contents

85.0 Calibration & Adjustment5. 1 Checking & Zero AdjustmentAfter installation on a system, the Model 730 should be zeroed. Figure 7 shows th

Page 4 - 2.0 Mechanical Installation

96.0 Maintenance & TroubleshootingThere are no general maintenance requirements for the Model 730 other than periodic zero adjustment. If the unit

Page 5

7.0 SpecificationsPerformance Data: Accuracy 1 ± 0.5% of Reading Optional ± 0.25% of Reading Resolution: Infinite. lmite

Page 6 - with 5-pin Terminal Strip

118.0 Returning the Model 730 for Repair Please contact a Setra application engineer (800-257-3872, 978-263-1400) before returning unit for repair to

Page 7 - 0.5” and 0.25” OD Tube

12 Calibration ServicesSetra maintains a complete calibration facility that is traceable to the National Institute of Standards & Technology (NIS

Page 8 - 3.0 Electrical Installation

159 Swanson Road, Boxborough, MA 01719-1304 Tel: 800-257-3872/978-263-1400, Fax 978-264-0292 Email: [email protected], Web: www.setra.comSS730 Rev.

Page 9

IndustrialHVACTest & MeasurementBarometricUltra High Purity/SanitaryVacuumSetra offers a complete line ofproducts for these industries:

Page 10 - 4.0 OPERATION

Contents1.0 Introduction ...12.0 Mechanical Installation

Page 11

1.0 Introduction Setra’s Model 730 capacitance manometer is a temperature compensated, absolute pressure transducer designed for accurate and repeata

Page 12

immediately, then replace the protective flange cover and store in an area where the temperature range is controlled between -20 to +85°C. The Model

Page 13 - 7.0 Specifications

SpanZero3in.mmFigure 2: Outline Drawing of Model 730 with 5-pin Terminal Strip, and 0.5” and 0.25” OD Tube. 1.50383.0577Mating ConnectorPin IinmmModel

Page 14

4Figure 3: Outline Drawing of Model 730 with 6-inch cable and15-pin D-sub Terminal Strip, 0.5” and 0.25” OD Tube.Pin 1Pin 15Model 730Model 730Model 73

Page 15 - Calibration Services

5 3.0 Electrical Installation The model 730 operates from a 12-30 VDC regulated power supply for the 0-10 VDC output version or from a 9-30 VDC regu

Page 16 - SS730 Rev.A 07/29/2008

Pin 10.123.051.5038.10Pin 9SpanZeroin.mm Pin Location Function 1 +Signal Output 9 Power Supply Common 4 Power Supply + VDC 8 Signal Output C