Setra-system Model 227 Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Capteurs Setra-system Model 227. Setra System Model 227 User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 8
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
Setra Model 227
Ultra High Purity 1 1/8’’ Down-Mount
C Seal and W Seal Pressure Transducer
1.0 GENERAL INFORMATION
Every Model 227 has been tested and calibrated before shipment. Model 227
performance specifications are listed on Page 6 of this Guide.
Setra Systems 227 pressure transducers measure gauge, compound or absolute
pressure and convert this pressure to a proportional high level analog output.
Two output versions are offered: A voltage output of 5 VDC FSO (Full Scale Output)
and 10 VDC FSO), and a current output of 4 to 20 mA .
Note: These instructions are also available on our website at www.setra.com.
1.1 EMC Certification
This product complies with EN61326 Electrical Equipment for Measurement,
Control and Laboratory use – EMC Requirements for Minimum Requirements
and Industrial Locations. Special caution should be taken to meet Standard
EN61000-4-5: 1995 Surge Immunity if any of the following conditions apply to
the installation: The product is installed outside; all or any part of the cable is
exposed to the outside; the cable is greater than 30 meters in length. In order
to meet the Surge Immunity requirements, the following conditions must be
followed during installation:
1. Shielded cable must be used, and the shield must be tied to earth ground (not power
supply ground) on at least one end of the cable shield/drain wire. The shield must be
maintained all the way from sensor to the power supply.
2. If unshielded cable is used, an earth grounded metal conduit fitting can be used to
replace the shielded cable.
3. For a sensor with a metal body or enclosure, the body/enclosure must be grounded to
earth. If a protective metal housing is used, the metal housing should be grounded to earth
4. If a protective plastic housing is used, the housing must be able to withstand at least
2 KV from the housing to earth ground, without damaging the circuit.
2.0 MECHANICAL INSTALLATION
2.1 Media Compatibility
Model 227 transducers are designed to be used with any gas or liquid compat-
ible with 316L Stainless Steel. Never submerge the transducers in any liquid.
2.2 Environment
The operating temperature limits of the 227 are as follows:
Operating Temperature Range °F (°C) -40 to +185 (-40 to +85)
Compensated Temperature Range °F (°C) +15 to +150 (-9 to +65)
Operating temperature limits with Option N1 are as follows:
Operating Temperature Range °F (°C) -22 to +176 (-30 to +80)
Compensated Temperature Range°F (°C) +15 to +150 (-9 to +65)
2.3 Pressure Fittings
Mounting - Model 227 pressure transducers are supplied
with 1.125 down-mount C or W seal base. (Note: The
torque values listed are typical for most systems, please
adjust the values to fit your system.) The tightening
sequence forms a leak tight seal between the base block
and the unit. Using 4 screws to install the unit, follow the
torque sequence pattern in Diagram 1. Use a torque driver and
torque all 4 screws to 25 in. lbs, repeat sequence and torque all 4 screws to 35 in. lbs,
then repeat sequence for 45 in. lbs.
2
Leak
check
hole
3
4
1
Diagram 1
Vue de la page 0
1 2 3 4 5 6 7 8

Résumé du contenu

Page 1 - 2.0 MECHANICAL INSTALLATION

Setra Model 227Ultra High Purity 1 1/8’’ Down-Mount“C” Seal and “W” Seal Pressure Transducer1.0 GENERAL INFORMATIONEvery Model 227 has been tested and

Page 2 - 3.0 ELECTRICAL INSTALLATION

23.0 ELECTRICAL INSTALLATION3.1 Voltage Output UnitsThe Model 227 voltage output transducer is supplied with a 6ft. multiconductorcable, Bayonet styl

Page 3 - Diagram 4

SetraTransducer(4-20mA)PowerSupplyLoad(Monitor)++_+__Diagram 43Current must flow in one direction only - Please observe polarity. (See Diagram 4) Wes

Page 4 - Diagram 5

should be adjusted with the unit at atmospheric pressure and theoutput adjusted as noted in Diagrams 5 & 6 on pages 4 & 5. Zero and spanadjus

Page 5 - Diagram 6

See Diagram 6 below for nominal “installed” output at atmospheric pressure forcompound pressure ranges.4.4 Current Output Span AdjustmentSpan or full

Page 6

5.0 MODEL 227 PERFORMANCE SPECIFICATIONSAccuracy RSS* (at constant temperature.) ±1.0% Reading±0.25% FSNon-Linearity, BFSL ±0.15% FSHysteresis 0.20% F

Page 7 - Setra Systems, Inc

Setra Systems, Inc.159 Swanson roadBoxborough, MA 01719Repair Dept. Fax #: 978-266-2158/Phone #: 978-266-2194✮ All lines must be filled out.✮ This for

Page 8 - SS2082 Rev.D 08/24/04

159 Swanson Road, Boxborough, MA 01719-1304Tel: 800-257-3872/978-263-1400, Fax 978-264-0292Email: [email protected], Web: www.setra.com SS2082 Rev.D 08

Commentaires sur ces manuels

Pas de commentaire